NanoFocus_μSurf

Introduction

Instructor: Andreas Best

Caution

[!IMPORTANT]

Doing Experiments

### Turning System ON - Turn on main switch in the cabinet - Turn on the light - Enter USER and Pass - Open μsoft application - If the master control has just been switched on, "Set reference" request should always be answered with “Yes”. - The reference position is the home position of the sample stage (x- and y-axis) and the linear axis of the sensor (z-axis). The coordinates of the home position are x = y = z = 0. Move it to most bottom and left.

[!CAUTION] - By setting the reference position a software stop is initialized, because the software doesn’t accept negative coordinates. In case of an accurate setting of the reference position, an accidental collision of the objective and the x/y-stage can be prevented.

### Doing a measurement

### Reference Calibration In standard microscope mode (see ‘Doing a measurement’, p. 17) the focus plane, i.e. a sharp picture of the sample surface, should be found by moving the measuring head up and down by means of the joystick. Then switch over to confocal mode, i.e. click the appropriate button in the software. The Nipkow disk is now situated in the light path. Look for the brightest area while moving through the confocal levels and adjust the lighting. Then search for the upper and lower limits of the measurement range and enter these data. As desired, enter some further measurement parameters and start the measurement.

General Information